首頁>產品中心>日本光明北川式氣體檢測管>氣體快速檢測管>281S氧/二氧化碳O2,CO2氣體檢測管281S OXYGEN CARBON DIOXIDE 2-10%

氧/二氧化碳O2,CO2氣體檢測管281S OXYGEN CARBON DIOXIDE 2-10%

簡要描述:

氧/二氧化碳O2,CO2氣體檢測管281S OXYGEN CARBON DIOXIDE 2-10%(O2),1-20%(CO2)必發生化是一家集研發、生產、銷售為一體的各種比長式氣體檢測管專業生產廠家,主要有氣體檢測管、一氧化碳檢測管、硫化氫檢測管、磷化氫檢測管、氨氣檢測管等產品。產品廣泛應用于煤炭、冶金、倉儲、建材、化工、焦化、環保等行業。www.r-biopharm.com.cn

更新時間:2025-09-03

發郵件給我們:554328895@qq.com

分享到: 1
在線留言
氧/二氧化碳O2,CO2氣體檢測管281S OXYGEN CARBON DIOXIDE 2-10%

氧/二氧化碳O2,CO2氣體檢測管281S OXYGEN CARBON DIOXIDE 2-10%(O2),1-20%(CO2)


1. PERFORMANCE
1) Measuring range:
Oxygen 2-10 %
Carbon dioxide 1-20 %
Number of pump strokes 1(100ml)
2) Sampling time :8 minutes/1 pump stroke
3) Detectable limit :0.001 %
4) Shelf life :1.5 years
5) Operating temperature :0~40℃
6) Temperature compensation:
Oxygen Necessary(See "Temperature CorrectionTable")
Carbon dioxide No temperature correction is necessary.
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change:
Oxygen White→Brown
Carbon dioxide Pink→Yellow
2. RELATIVE STANDARD DEVIATION
Oxygen RSD-low:10% RSD-mid.:10% RSD-high:10%
Carbon dioxide RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
Oxygen:Oxygen reacts with alkaline pyrogallol.
Carbon dioxide:CO2+2KOH→K2CO3+H2O
4. CALIBRATION OF THE TUBE
Oxygen?Carbon dioxide STANDARD GAS CYLINDER METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
Ordinary combustion gases do not affect the reading value and discolouration.
5. CONCENTRATION CORRECTION
On account of interaction of indication between carbon dioxide and oxygen, the true concentration is calculated
by the following equations.
[EXANPLE]
At the measuring temperature of 30℃, when oxygen indication is 5%(temperature corrected value 5.3%)and
carbon dioxide indication is 10%, each true concentration is calculated as bellow.
Accordingly true concentration :
Carbon dioxide 9.47%
Oxygen 4.77%
OXYGEN CARBON DIOXIDE DETECTOR TUBE XYGEN DETECTOR TUBE
23456789
10
4
2 1
6
8
10
12
14
16
18
20
%
GAS To pump
Tube No.
281S OXYGEN?CARBON DIOXIDE
TEMPERATURE CORRECTION TABLE
Scale True Oxygen Concentration (%)
Readings 10℃ 20℃ 30℃ 40℃
(%) (50°F) (68°F) (86°F) (104°F)
10.0 9.7 10.0 10.3 10.5
9.0 8.7 9.0 9.3 9.5
8.0 7.7 8.0 8.3 8.5
7.0 6.7 7.0 7.3 7.5
6.0 5.7 6.0 6.3 6.5
5.0 4.7 5.0 5.3 5.5
4.0 3.9 4.0 4.2 4.5
3.0 2.9 3.0 3.1 3.3
2.0 2.0 2.0 2.0 2.1
CO2 %=(CO2)- 100
(O2)×(CO2) equation(1)
O2 % =(O2) - 100
(O2)×(CO2) equation(2)
CO2 %= 10 - = 10 - 0.53 = 9.47
= 5.3 - 0.53 = 4.77
100
5.3 × 10
5.3 × 10
O2 % = 5.3 - 100
Where
(CO2):indication of carbon(%)
(O2) :corrected oxygen concentration(%)